Ventilators Manufacture

From ATI Chennai IT and ITES Wiki

Jump to: navigation, search

Contents

Ventilators

Ventilators are needed to keep respiratory conditions at bay and is so for Corona Virus patients. Original PDF Document - 512 kb, 25 pages

Ventilator Specifications - RMVS

Medicines and Healthcare Products Regulatory Agency (MHRA) is the UK body that approves medical equipment for use.

  • 100% reliable for 14 days continuous operation
  • Small and light enough to fix to a hospital bed
  • Robust enough to survive falling from bed to floor
  • Sense when a patient stops breathing and switch from the assisted breathing mode to a mandatory setting
  • Connect to hospital gas supplies
  • Have least 20 minutes of backup battery
  • Batteries should be swappable
  • All parts detachable for manual disinfection
  • Fault Alert like lack of oxygen supply
  • Oxygen percentage and other monitoring data on dashboard display
  • Low learning curve typically less than 30 minutes
  • Support a range of 10 to 30 breaths per minute, rising in increments of 2
  • Adjustable inhalation to exhalation duty cycle
  • Tidal volume (volume of single breath) should be a minimum of 6 ml / Kg of body weight - ie., 500 ml for 80 Kg person
  • Adjustable settings from 250 ml to 800 ml in increments of 50 ml
  • Adjustable Oxygen percentage from 30% to 100% in increments of 10%

Open Source initiatives

Hardware

ARM based Options